+ 91 4065546999 || + 91 9100971894
Products
RI detectors. For detection of UV inactive substances(e.g. sugars, polymers etc.) in HPLC or FLASH chromatography applications.
Analytical version up to flow rate 5 ml/min.
Differential Refractive Index Detectors are used for detecting little or no UV active samples such as sugars or polymers. The Shodex RI-101, employs a number of improvements for shortening the time for stabilization. By using a micro-processor it is capable of automatic start-up from eluent replacement to checking the base-line stabilization. The detector is fitted with external input and output terminals as well as RS232C communication ports, permitting advanced automation by linking the detector with various HPLC systems, enabling advanced automation by combining the detector with various chromatographs.
We also offer preparative (RI-102) and semipreparative (RI-104) version of this detector - see download files box.
Refractive index range | 1/4 – 512 μRIU |
Ordinary flow rate | 0,2 to 3 mL/min |
Maximal flow rate | 10 ml/min |
Noise level | 2.5nRIU (response 1.5 sec) |
Linearity range | 600 μRIU |
Cell volume | 8 μl |
Dead volume | Total: ca.670 uL |
Solvent contact parts | SS 316, Teflon, Quartz glass |
Auto zero resolution | < 1 (@ 8mV/u RIU) /4 (@ 2mV/uRIU) n RIU |
Signal output | 0 – 1V |
Adjusted temperature | OFF, 30-50°C(1°C increment ), 77°C Temp. Fuse |
Power input | 150 VA |
Dimensions (W x H x D) | 260 x 400 x 200 mm |
Weight | 13 kg |
Power supply | 100-230 V ± 10% , 50, 60 Hz |